A13914AF
8-10
User Defined Reagents
Processing Parameters
Figure 8.1
Timing Chart
Third Inject: Component
This designates the compartment of the cartridge which will be pipetted third. If Third Inject is not
used, the default compartment selection is
None
and should not be modified.
TIME (SECO
N
DS)
Reagent Blank Read Window
Reaction Read Windows
E011407L.EPS
C
u
v
ette
W
ash
Completed
Pr
imar
y
R
eagent
Fill
Sample
Inject
C
u
v
ette
W
ash
T = -1
8
0
T = -172
T = 0
T = 9
T = 73
8
3 to 6 min
u
tes
DxC 600/
8
00 Timing Chart
Second Inject Add Time (T = -1
8
0, 9 to 73
8
)
Third Inject Add Time (T = -172, 9 to 73
8
)
Table 8.19
Options for the Third Inject Component
None, A, B, or C