A13914AF
12-22
Troubleshooting Calibration and Result Errors
Error Code – Definitions
Slope Offset Adjustment —
The test result associated with this Result Error code was calculated with a
modified calibration. Slope and/or Offset were modified and the result was directly affected by these
calibratrion adjustments. Refer to
in CHAPTER 4,
Reagent Load/Calibration
in this
manual for additional information
Substrate Depletion —
The difference (delta) between the initial absorbance taken after sample or trigger
reagent inject and the final absorbance data within the reaction read window exceeds specifications.
Rerun sample with ORDAC enabled (if appropriate) or dilute sample and rerun..
Temperature —
The operating temperature of the system is beyond 0.1°C from the set point value. All
cartridge chemistries will be reported with a flag.
Test Aborted —
Scheduled test has been aborted for non-specified reason. This usually occurs when a
system error compromises chemistry testing or sytem has been stopped while running.
Under Range Detection and Correction (URDAC) High/Low —
A reaction exceeded the range of URDAC
values that the instrument will report.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......