PCR-258

OMNI Tip™

HOMOGENIZING KITS

USER MANUAL

Summary of Contents for PCR-258

Page 1: ...REV B PCR 258 OMNI Tip HOMOGENIZING KITS USER MANUAL...

Page 2: ...ithout charge For the balance of the period parts will be provided but service will be charged at established labor rates This warranty does not include normal wear from use it does not apply to any i...

Page 3: ...ion fire or injury Use this product only for its intended purpose as described in this booklet DO NOT use attachments not recommended by the manufacturer DO NOT operate the product if it is damaged in...

Page 4: ...for those applications WARNING Keep all housings in place and in working order WARNING Remove all tools from the generator probe before turning the motor on WARNING DO NOT use the motor in a dangerou...

Page 5: ...he Omni TH 3 1 Stand Mounting the Omni TH 9 3 2 Installing Rotor Stator Generator Probes 9 3 3 Omni Tip Plastic Probes 10 3 4 Omni TH Operation 10 3 5 Theory of Operation 11 Section 4 Operational Guid...

Page 6: ...n for more information Omni s Tissue Homogenizing Kits are designed for situations where molecular contamination between samples cannot be tolerated Stainless steel generator probes are available for...

Page 7: ...ssue Homogenizing Kits is required for mounting Omni Tips to the Omni TH Homogenizer and is not disposable See Section 5 for more information about Omni Tips Part Number A1000SB 12 120 30750 30750H 32...

Page 8: ...03 100 mL Height motor only 22 9 cm 9 in Weight motor only 466 g 16 4 oz Electrical Requirements 115 V 60 Hz or 220 V 50 Hz Standards Approval Compliance 220V CE Certified Description Motor Drive Unit...

Page 9: ...n firmly The unit will lock in place 4 To remove the TH press the black button on the back of the opening of the post clamp assembly and lift the motor from the cradle 3 2 INSTALLING ROTOR STATOR GENE...

Page 10: ...he sample vessel can significantly affect flow patterns within the vessel This also affects processing efficiency As a rule of thumb the generator probe usually operates most efficiently at a depth of...

Page 11: ...eraction of the rotor knife with these windows sets up a shearing action reducing the particle size of the sample The speed differential between the rapidly moving portion and the relatively stationar...

Page 12: ...n times for repeated use they are not recommended for extended use with phenol or guanidine based reagents see Section 5 3 for chemi cal resistance compatibility Disposal may be required when used in...

Page 13: ...s Processing Speed Omni Tip Homogenizing Kits include a handheld Omni TH Homogenizer with motor speeds up to 35 000rpm Operating at the top speed enhances processing efficiency of the generator probes...

Page 14: ...ntamination for many applica tions the probe can be cleaned sterilized or disinfected and reused by the procedures described below NOTE Omni Tip plastic probes are not factory presterilized or disinfe...

Page 15: ...euse However phenol or guanidine based reagents commonly used in PCR studies are not recommended for extended use with Omni Tips Exposure to these chemicals shortens the life of Omni Tips and usually...

Page 16: ...ft begins to warp as a result of autoclaving discard and lower the cycle time or temperature level accordingly for future runs 5 5 CLEANING THE OMNI TIP ADAPTER The Omni Tip 12 120 and A1000SB adapter...

Page 17: ...he Omni TH is plugged in and turned on but is not functioning Check that the brushes are not worn and are correctly installed Replace if necessary see Section 3 2 Motor is turned ON and makes a makes...

Page 18: ...18 NOTES...

Page 19: ...19 NOTES...