A13914AF
9-55
Maintenance
Six-Month Maintenance
9
3
Prevent leakage of reagent from the flow cell.
•
Select:
—
Utils
from the menu bar,
—
2 Maintenance
,
—
3 ISE Service
,
—
Continue
.
The flow cell is prepared for electrode maintenance.
4
Refer to the appropriate electrode procedures for servicing other electrodes (
).
•
Chloride Electrode (or
Replace the Chloride Electrode Tip
)
•
CO
2
Measuring Electrode (
•
Sodium Electrode (
Replace the Sodium Measure/Reference Electrode
).
Removal and Installation of the Electrodes
CAUTION
To prevent damage due to electrical static discharge (ESD), wear the wrist ground
strap when you perform this procedure. For further instructions, refer to
in the beginning of this chapter.
1
Disconnect the electrode cable from potassium connector (1) and calcium connector (2).
Figure 9.36