A13914AF
9-18
Maintenance
Monthly Maintenance
Figure 9.7
DxC 800 ISE Module
3
Unscrew and remove the bottle cap from the alkaline buffer reagent. Hold the cap with attached
straw over a container that catchs the residual fluid from the alkaline buffer return line #33.
4
Prime the line 10 times.
•
Select:
—
Utils
from the menu bar,
—
1 Prime
,
—
MC F4
,
—
ISE CO
2
Alkaline Buffer
.
•
Type
10
in the
Number of primes to repeat
field.
•
Select
Start Prime
. Reagent line is drained of old reagent.
5
Remove reagent input line #30 (1) (line draining into damper) from the top of the in-line filter.