A13914AF
6-7
Sample Programming and Processing
Sample Programming and Processing
6
Manually Assigning Rack and Position
It is only necessary to manually assign a rack and position if:
•
The system is set up in bar code mode and a sample will not be bar coded,
OR
•
A bar code is unreadable by the system.
1
Select
Samples
from the menu bar.
2
Select
Racks F8
.
3
Select a rack with an open position.
OR
Type the Rack number.
4
Type the desired manual assignment into the selected rack/position.
5
Select
OK
to accept the assignment.
OR
Select
Cancel
to leave assignment screen without saving assignments.
6
Select
Program F10
to return to
Sample Program
screen.
Processing Samples Manually
Manual programming is used when there is no LIS or the LIS is not available. Samples may or may
not be bar coded. The sample is identified by a Sample ID and/or rack and position. The user selects
tests for programming.
The following items require manual programming:
•
Samples without bar codes and no sample programming.
•
Samples with bar codes and no sample programming.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......