A13914AF
9-126
Maintenance
As-Needed/As-Required Maintenance
2
Open the left door on the DxC. The ISE drain pump is located on the inside left wall.
CAUTION
BIOHAZARDOUS MATERIALS HAZARD. Observe all biohazard precautions when
you perform maintenance, service, or troubleshoot the system. This includes, but
may not be limited to, wearing gloves and eye shields and washing hands after
working on contaminated portions of the system.
MOVING PARTS HAZARD. Do not place hands near any moving parts while the
system is operating.
3
Refer to
.
•
Lower the clear plastic cover (1).
•
Disconnect both pump tube fittings (2 and 3), and move the tube out of both pinch clamps
(4 and 5).
•
Manually rotate the pump roller mechanism (6) approiximately half a turn while removing
the tube. Use caution as the tube may contain residual fluid.
Figure 9.88
ISE Drain Pump
4
Discard the used tube in a bio-hazardous waste container.
1.
Plastic Cover
2.
Pump Tube Fitting
3.
Pump Tube Fitting
4.
Pinch Clamp
5.
Pinch Clamp
6.
Pump Roller Mechanism
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......