A13914AF
4-25
Reagent Load/Calibration
Chemistry Bypass
4
Chemistry Bypass
Introduction
In the event of a calibration failure for a chemistry which has been previously sample programmed,
Chemistry Bypass permits the continuation of analysis of all other programmed tests without
having to deprogram a failed chemistry. The chemistry can be bypassed while the system status is
Standby, Disabled
or
Running
.
•
The system will not aspirate sample or dispense reagents for bypassed chemistries. They will
not be marked as
Pending
.
•
Bypassed chemistries are logged on the Post-Run Summary Report.
•
Bypassed chemistries must be calibrated to process samples.
Bypass a Chemistry
1
Select
Rgts/Cal
from the menu bar.
2
Select the Chem position to bypass.
3
Select
Options F6
.
4
Select
Bypass chemistry
from the pull-down menu.
Remove a Bypassed Status
Recalibrate the chemistry to remove a bypassed status.