A13914AF
9-92
Maintenance
As-Needed/As-Required Maintenance
5
Remove line #23 from the bottom of the flow cell. Attach the connector fitting (1) on the small
diameter tubing, included in the Flow cell flush kit, to the syringe. Connect the open end of the
tubing to the bottom of the flow cell.
Figure 9.70
6
Without removing the hemostat, remove line #35 from the top of the flow cell. Remove line #27
from the front of the flow cell in order to allow residual acid to leak out.
7
Draw back on the syringe. This withdraws any remaining solution left in the flow cell and
removes any material that can be causing the occlusion of the small lumen.
NOTE
The procedure may be concluded here if the problem has been resolved. Proceed to Step 10.
If the problem is not resolved, proceed to the next step.
8
Re-attach line #27 to the front of the flow cell. Remove the syringe and empty the contents. Fill
the syringe with ISE Buffer.