A13914AF
5-9
Quality Control
Edit a Control Definition
5
4
Select the number next to the
Control Name
field to select from predefined Control Names. If
necessary, use the scroll bar to view additional Control Names.
OR
Type the number in the
No.
field.
The
Program Control
screen opens. Make sure the Control Lot number on the screen is the same
as the lot number on the Control bottle.
5
Type the rack/position number if the bar code is not available.
6
Select chemistries to be run.
NOTE
In Manual QC Programming, the system automatically runs the oldest cartridge onboard for the
selected chemistry. Specific cartridges (other than default) may also be selected. To access duplicate
or specific cartridges, select
Rgt Cart F8
. Then select the desired cartridges and select
OK
.
7
Select
Save F10
to save the control programming and return to the
Program Sample
screen.
IMPORTANT
If all tests programmed for the control do not complete testing (for example: calibration
required, cal time exceeded or disabled), the next time the control ID is loaded, only the incomplete tests
will run. To run all chemistries (autogeneration of control), the control ID must be cleared manually before
reloading the sample.
Edit a Control Definition
Introduction
The assigned mean, SD, constituent code, and bar code IDs of a previously defined control may be
edited. The QC File Number associated with any defined control may not be edited.
Edit Precautions
•
Changing the mean and/or standard deviation may affect subsequent QC statistical data.
Changing previously defined mean and/or SD to zero will set the mean to zero, and the SD to
99999.
•
Subsequent data points will be compared to the new mean and SD.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......