A13914AF
9-77
Maintenance
As-Needed/As-Required Maintenance
9
3
Center the membrane retainer (1) over the membrane clamp (3). Press the membrane retainer
into the clamp, thus securing the membrane (2).
Figure 9.57
4
Place the retainer assembly upside down on the work surface so the clamp is on top and retainer
on the bottom. Grasp the electrode carefully and press it firmly against the membrane and into
the membrane clamp.
Figure 9.58