A13914AF
5-23
Quality Control
Archive QC
5
Archive QC
Introduction
The Archive QC function archives control definition (control name, lot number, sample type, QC File
number, QC Log selected chemistries, assigned mean and SD, constituent code, and cumulative
mean, SD and N) and results to a diskette. Archiving is available from the hard disk only. The
archived diskette can be used to review data, but not to modify QC files. The system must be in
Standby
to archive QC data.
The QC database can store at least 35,000 results. If a new result is processed and the QC Results
database is full, the oldest result related to the new result will be overwritten. If the QC Results
database is full and no results are associated with the QC File number of the new result in process,
the new result will not be saved. The operator will be notified that the new result was not saved.
When this dialog box notification appears, it may be appropriate to archive.
Archive QC
IMPORTANT
To archive Current QC Data to a diskette, the diskette must be Double-sided, High Density.
1
Select
QC
from the menu bar.
2
From the
QC
screen, select
Archive F8
.
3
Select
1 Archive to Disk
.
OR
Type
1
in the
Option No.
field and press
(
Enter
)
.
4
At the prompt, type an operator ID of 1 to 3 alphanumeric characters.
5
To confirm the continuation of the archive process, select
OK
.
To discontinue process, select
Cancel
.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......