A13914AF
3-33
System Setup Options
Auto Generation of Control
3
Auto Generation of Control
Introduction
When this feature is enabled, if a sample is loaded with a defined control ID, the instrument will
automatically
run any chemistry that is on-board and runnable for that control.
When Auto Generation of Control is enabled, the automatic Multiple Cartridge option is available.
With the Multiple Cartridge option enabled, the system automatically runs all cartridges on-board
and runnable for any chemistries for an Auto Generated control. Enabling Multiple Cartridge
without Auto Generation has no effect on the manual programming of control samples.
Setup
From the
Setup
screen, scroll down and select
29 Auto Generation of Control
to enable/disable either
feature.
The system must be in
Standby
,
Stopped, Startup, Instrument Down, or Homing
to make modifications.
The default is
disabled
.