A13914AF
5-12
Quality Control
Review a Control Definition
Review a Control Definition
Introduction
A control definition can be reviewed from either current QC data or archived QC data. Archived QC
data may not be modified, only printed. Current QC data may be reviewed, deleted, or printed.
Review a QC
1
Select
QC
from the menu bar.
2
Select a control number.
3
Select
Review F1
.
All current information and selected chemistries appear, including serum indices.
•
Control ID F2
can be accessed to view the eight bar code IDs.
•
Delete F3
can be used to clear a control.
•
Select
Print F9
to print a control.
•
When reviewing QC definition, no cursor will be available because changes are not allowed.
4
Exit the
Review
screen by selecting
Done F10
.
Delete a Control
Introduction
Previously defined controls may be deleted and removed from the system. Clearing QC ranges and
applicable data can be done for Current QC data only. Controls may also be deleted through define,
edit, and review QC functions.
Controls with
Incomplete
status cannot be deleted. Clear the Control ID at the
Program Sample
screen.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......