A13914AF
10-16
System Status and Commands
Enable/Disable Modules
2
Select
5 Pause Waste B
.
Waste B removal will be paused until Resume Waste B is selected. The
Instrument Commands
screen
remains so that Resume may be selected. Pause Waste B will be gray, to indicate it is already paused.
Resume Waste B
1
Select
Instr Cmd
from the menu bar.
2
Select
6 Resume Waste B
.
The system will return to its normal operational state. The
Instrument Commands
screen closes
automatically.
Enable/Disable Modules
Introduction
Enable/Disable allows the user to Enable and Disable an individual subsystem module, the MC side,
or the CC side. This function allows some subsystems to operate while others are disabled due to
error conditions.
Additional Information
The user can manually enable or disable the following items:
•
One or more individual MC modules (cups and ISE)
•
All of the MC side
•
Photometer
•
LPIA Module
•
All of the CC side
•
Lower Reagent Carousel (1–30)
Notes:
The optional LPIA Module must be installed to be able to enable or disable the LPIA Module.
A module should be enabled or disabled only when the system is in
Standby
or is
Stopped
.
To complete the enable for 1-Blade CTS, or LPIA, the operator must stop and home the DxC.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......