A13914AF
11-6
Utilities
Maintenance
2
Select the desired cup(s) and specify Reagent and/or DI water.
AND/OR
Select a single ISE option by choosing a check box.
AND/OR
Select the
Sample Delivery Subsystem
check box.
3
Type the desired number of primes (1–99) in the
Number of primes to repeat
field.
4
Select
Start Prime
. The system status changes to
Priming
and the number of primes remaining
appears.
5
Upon completion of the requested prime, the number of primes remaining will be blank and the
system status returns to
Standby
.
Maintenance
Maintenance is routine instrument cleaning and replacement of parts as they become worn. Proper
maintenance is essential for optimal instrument performance.
Additional Information
Refer to the
in this manual for maintenance instructions.
Event Log
Introduction
The Event Log is a record of events and errors that are logged (for example, CAU: DI Water Canister
Level Low. Message ID 0x01FA000E.). This logged information can be used as a troubleshooting tool
if a problem is encountered while operating the instrument.
The entire Event Log can be viewed, printed, or copied to disk. It is possible to select only a portion
of the Event Log to work with by specifying classes and/or a date and time range.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......