0x02620019
2
CC Sample Delivery
Subsystem Motion
Error, Pausing
CAU:
CC Sample Delivery
Subsystem Motion Error,
Pausing
Message ID:
0x02620019
Subsystem:
ccScheduler
This occurs after two motion errors on
the Sample Delivery Subsystem. Level
Sense errors will not trigger this error.
The system goes into
Pausing
and then
Stopped
state.
•
Raise canopy and check for obvious obstruction
of the sample probe or syringe.
•
the system.
While system is homing, observe motion of the
CC sample syringe and CC Sample probe. The
probe and syringe should complete their
operations in a smooth manner.
•
If the error continues, check the event log to
determine if the syringe or probe is responsible
for the error.
•
If the CC sample probe is responsible, perform
the appropriate vertical or horizontal alignment.
Refer to Step 3 in the
procedure.
•
If the Syringe is responsible, check for proper
seating of syringe assembly.
•
Contact Beckman Coulter Support Center for
additional information.
0x0262001A
4
CC Sample Probe
Obstruction
Sample ID:
Rack ID:
Rack Position:
This occurs on first
attempt
CAU:
CC Sample Probe
Obstruction
Message ID:
0x0262001A
Subsystem:
RASched
Sample ID:
Rack ID:
Rack Position:
Sample Error Count:
Probe Error Count:
Cycle Number:
This error occurs when the CC sample
probe ODC (Obstruction Detection &
Correction) indicates that the probe is
obstructed during sample aspiration.
•
No operator action is necessary.
•
The system will attempt a second aspiration
from this sample.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......