A13914AF
1-52
System Description
Cartridge Chemistry: Calibration Theory
Model #9
Math Model #9 is an extension to model #1, the four-parameter log-logit function.
The "c" is allowed to be 1 or -1.
If c = +1, then this is equivalent to model #1.
If c = -1, an alternative function is being used.
This function cannot be solved directly for concentration. The instrument uses an
iterative method to determine the sample value.
Polynomial
Exponential
(PXP)
Quadratic
(POLY2)
Lorentz
Double Inflection
Model DP4
Table 1.18
Math Models for Non-Linear Chemistries
(Continued)
Type
Formula
E014476L.EPS
E014477L.EPS
E014478L.EPS
E011537L.EPS
R =
K
c1
+
K
c2
1 +
a
1
1
+
a
2
conc
conc
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......