A13914AF
8-20
User Defined Reagents
Expanded User Defined Chemistry Feature
Expanded User Defined Chemistry Feature
Introduction
The Expanded User Defined Chemistry (or UDR for User Defined Reagent) feature adds any or all of
the following options for up to 10 UDR chemistries:
•
Retain calibration data when a cartridge is removed from the instrument.
•
ORDAC capability.
•
Onboard dilution of samples.
•
Selection of 940 nm as the primary wavelength.
The options are enabled when
UDR+ F3
is selected from the
Define/Edit User Defined Chemistry
screen. An Expanded UDR chemistry is denoted by a “+” placed in front of the UDR name when
viewed in the
User Defined Chemistries
screen.
Cal Save
Cal Save allows the calibration parameters to be retained when the cartridge is removed from the
instrument. When the cartridge is reloaded within the calibration time limit, recalibration is not
required.
Enable/Disable Cal Save
1
The instrument must be in
Standby
or
Stopped
mode.
2
Select
Setup
from the menu bar.
3
Select
15 User Defined Chemistries
.
4
Select the number to be defined.
OR
Type a number in the
Option Number
field.
5
Select
Define F1
.
6
Select
UDR+ F3
.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......