A13914AF
3-17
System Setup Options
Reportable Ranges Setup
3
Setup
From the
Setup
screen, select
8 Report Setup
to make selections or changes to report setup
specifications.
Report Setup may be viewed at any time.
The system must be in
Standby
,
Stopped, Startup, Instrument Down, or Homing
to edit Patient or
Control Report formats, Print Option, or the Inter-Laboratory information.
Reportable Ranges Setup
Introduction
The three types of range limits that can be applied to each analyte are:
•
Analytical Range
•
Instrument Printable Range
•
Reportable Range
Analytical Range
The Analytical Range is an internal system limit verified by Beckman Coulter.
Refer to the Synchron Clinical Systems
Chemistry Information Manual
and the Synchron Clinical
Systems
Chemistry Reference Manual
for Analytical Ranges by analyte.
Instrument Printable Range
The Instrument Printable Range is an internal system limit, specified in the Reportable Range
Setup
screen. The Instrument Printable range exceeds the Analytical Ranges by a factor. This allows for
the printing/reporting of a result slightly above or below the Analytical Range, given small
precision variations. A result outside of the Instrument Printable Range is suppressed. The
suppressed results are flagged OIR HI or OIR LO (
O
ut of
I
nstrument
R
ange).
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......