Maintenance
As-Needed/As-Required Maintenance
9
9
Remove Line #35 from the top of the flow cell. Attach the connector fitting (1) of the large
diameter tubing, included in the Flow cell flush kit, to the syringe. Connect the open end of the
tubing to the top of the flow cell. Flush the buffer through.
Figure 9.71
10
Reattach the lines to the flow cell and remove the hemostat.
11
When all components are in place for the flow cell select
OK
to prime the flow cell. Select
Close
in the pop-up window to exit the
ISE Service
screen.
12
Prime the flow cell 5 times.
•
Select:
—
Utils
from the menu bar,
—
1 Prime
,
—
MC F4
,
—
ISE All
check box.
•
Type
5
in the
Number of primes to repeat
field.
•
Select
Start Prime
.
CAUTION
Pinch hazard. Keep fingers clear of the ISE module as you lower it.
13
Lower the ISE module by releasing the locking pin. Be careful NOT to pinch tubing as the module
is lowered.