A13914AF
3-27
System Setup Options
Disable Service Monitor
3
Setup
From the
Setup
screen, scroll down and select
17 Reserved Racks/Obstruction Detection
to enable/
disable CTS, enable/disable obstruction detection, assign reserved rack numbers for small sample
volume containers, HbA1c or IBCT, or to restore defaults.
The system must be in
Standby
,
Stopped, Startup, Instrument Down, or Homing
to make modifications.
When changing the CTS option, the operator must
Stop
and
Home
the system to completely enable
the feature.
Defaults:
•
CTS:
Off
•
Obstruction Detection:
On
•
Reserved Racks:
None Reserved
•
Reserved Racks for HbA1c or IBCT:
None Reserved
Disable Service Monitor
Introduction
The Disable Service Monitor option permits the service monitor to be disabled.
When the
Metering Counter
is activated, the
Service Monitor
must be disabled. Refer to the Synchron
Clinical Systems
Metered-Use Manual
for information regarding the metering option.
Setup
The system must be in
Standby
or Stopped
to change this option.
From the
Setup
screen, scroll down and select
18 Disable Service Monitor
.
The default is
Disabled
.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......