A13914AF
1-3
System Description
Operational Conditions
1
Notes on the System Power
The system can operate from any standard 3-wire electrical outlet and is wired as shipped from the
factory to operate on 220 VAC, 50/60 Hz.
IMPORTANT
Line Voltage from the electrical outlet should be free of spikes, fluctuations, and dropouts for
protection of the electronic circuitry.
CAUTION
Only operate the system from a 3-wire power source. DO NOT use a 2-prong
adapter or a 2-wire AC power source.
Environmental Conditions
Table 1.4
Power Requirements – DxC Console (PC System and Monitor)
Item
Requirement
Operating range
100–120 VAC ± 10% (90–132 VAC); 4A
OR
200–240 VAC ± 10% (180–264 VAC); 2A
Frequency
50/60 Hz
BTU generated
1,500 BTU/hour
Power connector
15 A current rating, IEC 320 standard connector
Table 1.5
Power Requirements – Okidata B4350 LED Printer
Item
Requirement
Operating range
110–127 VAC ± 10% (99–140 VAC); 3A
OR
220–240 VAC ± 10% (198–264 VAC); 1.6A
Frequency
50/60 Hz
BTU generated
1228 BTU/hour (printing); 34 BTU/hour (standby power save)
Power connector
15 A current rating, IEC 320 standard connector
Table 1.6
Temperature, Humidity and Elevation
Item
Specification
Environment
Indoor use only
Ambient temperature
+18°C to +32°C
Warm-up time
30 minutes (time to reach operating temperature)
Relative humidity
20–85% relative, non-condensing
Elevation
Up to 4,200 ft (1280 m)
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......