8-2
User Defined Reagents
Requirements and Precautions
User Defined Reagent Cartridges
The user-defined reagent is placed in generic, non-bar code labeled, three-compartment cartridges
that are loaded onto the system manually. These cartridges are available in a box of 12 (PN 442835).
The label on the cartridge should be marked with the reagent name and expiration date. The
maximum and minimum fill volumes, required to allow for accurate level sensing, are indicated in
the table below.
IMPORTANT
Since Beckman Coulter does not manufacture or otherwise control the reagents that may be
used in user defined reagent cartridges, Beckman Coulter makes no warranty whatsoever with respect
to such reagent's performance (including test results), their effect on the system or required system
maintenance or the frequency thereof, or their effect on operator safety. User assumes full responsibility
for use of the proper test protocol and test result generation for the reagent(s) selected by the user and
for any errors or omissions associated therewith. BECKMAN COULTER EXPRESSLY DISCLAIMS ALL
WARRANTIES WITH RESPECT TO THIS PRODUCT WHETHER EXPRESS OR IMPLIED, INCLUDING
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
IMPORTANT
Non-Beckman Coulter reagents, calibrators, and controls can contain components, not listed
on the insert, which may carry over into the system causing chemical or spectral interference. This
carryover could adversely affect results on a properly performing system. Manufacturers of user-defined
reagents should be contacted for disclosure of potentially interfering substances, such as preservatives.
Table 8.1
Fill Volumes for Accurate Level Sensing
Compartment A
Compartment B
Compartment C
Maximum Fill Volume
110 mL
18 mL
4 mL
Minimum Fill Volume
6 mL
1 mL
0.3 mL
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......