A13914AF
8-17
User Defined Reagents
Exit Check Criteria
8
Option 2
Prepare the reagent according to the manufacturer’s instructions. With a manual
spectrophotometer or automated instrument, obtain the delta absorbance per minute on at least
ten replicates of a normal or high normal sample at each of the appropriate wavelengths selected.
Average the delta absorbance per minute obtained at each wavelength. The extinction coefficient
at the secondary wavelength may be calculated by use of a factor or correction coefficient (R)
derived from the ratio of the delta absorbance per minute at the secondary and primary
wavelengths. The following formula may be used:
Exit Check Criteria
Introduction
When preparing to save a chemistry protocol after editing, the parameters are checked for
correctness, completeness, and consistency. Chemistry protocols that fail any Exit Check criteria
are not allowed to run. The following requirements specify the Exit Check criteria and parameters
presented for modification when the user elects to correct detected protocol errors.
Wavelengths
•
The primary wavelength must not be
None
.
•
The secondary wavelength must not be
None
.
•
The primary wavelength must be different from the secondary wavelength.
Calibration
•
The calibration time limit must be greater than zero hours if the number of calibrator levels is
greater than zero.
•
The sequence of calibrators must be in ascending target value order.
e
2
= e
1
× R
where:
e
2
=
Extinction coefficient at the secondary wavelength
e
1
=
Extinction coefficient at the primary wavelength
a
a. The extinction coefficient of the chromophore at the primary wavelength is usually specified by the reagent manufacturer.
E01442
8
L.EPS
R =
Delta absorbance / min
u
te at primary wavelength
Delta absorbance / min
u
te at secondary wavelength
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......