Maintenance
Electronic Maintenance Log
Four-Month Maintenance
•
Clean MC Reagent Lines
Six-Month Maintenance
•
Replace CALC Electrode Tip (6 months or 80,000 samples)
•
Replace K Electrode Tip
•
Replace Glucose Sensor
Electronic Maintenance Log
Feature Summary
Maintenance items are grouped together by frequency of performance. Only the maintenance items
appropriate for the particular system and installed hardware are shown. For example, a system that
does not have Closed Tube Sampling (CTS) hardware does not show the maintenance items for that
hardware. On the
Scheduled Maintenance Log
screen these groups have been placed under separate
tabs labeled:
•
Weekly
•
Monthly
•
Two-Month
•
Three-Month
•
Four-Month
•
Six-Month
To identify the maintenance procedures that are DUE, follow the yellow highlights as listed below:
•
Select
Utils
from the menu bar,
•
Select
2 Maintenance
,
•
Select
Log F2
,
•
Select the appropriate highlighted maintenance tab.
The yellow highlights appear according to the following schedule:
•
Weekly - 8 days
•
Monthly - 31 days
•
Two-Month - 61 days
•
Three-Month - 91 days
•
Four-Month - 121 days
•
Six-Month - 181 days
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......