A13914AF
9-104
Maintenance
As-Needed/As-Required Maintenance
5
Exit the Maintenance menu.
IMPORTANT
To resume autoprime and return the system to
Standby
, you must exit the Maintenance
menu.
6
Remove the CCWA cleaning solution cartridge.
•
Recap the cartridge and store at room temperature. Reuse at the next maintenance
interval.
•
Discard the cartridge after 180 days.
Clean the MC Sample Crane and Collar Assembly
Perform the MC Sample Crane and Collar Wash Assembly procedure whenever:
•
A build up of residue is observed on the collar wash assembly, or
•
It is suspected that a dirty collar wash assembly is causing increased level sense errors.
Materials Required:
•
One sample rack
•
One 2 mL cup
•
Undiluted household bleach
*
(Sodium hypochlorite 5.25%) (PN A32319)
1
Select:
•
Utils
from menu bar,
•
4 Diagnostics
,
•
Functional Area
1 Sample System
,
•
Subfunctional Area
1 MC Chemistry Sample Crane
,
•
Test Number
1 Level Sense Test
.
2
Type the Rack ID into the
Rack ID
field. Select the appropriate Rack Position. Select
Cup
as
container type.
3
Fill the cup with bleach. Place it in the specified rack and position, and then place the rack in
the autoloader.
4
Select
Start
.
The System automatically loads the rack and the MC probe aspirates the sample.
*
DO NOT use bleach that contains additives (for example, Ultra Bleach Advantage).
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......