A13914AF
2-5
Preparing Samples for Analysis
Preparing Samples for Analysis
2
Sample Tubes Validated for CTS
Capped tubes can be run directly on the system without removing the cap. The CTS assembly cuts
a small hole in the cap and the sample probe aspirates a sample directly from the tube. With this
procedure, samples are processed faster and more safely.
CTS Validated Sample Tubes
Only validated sample tubes can be run on UniCel DxC Systems with CTS. However, capped and
uncapped tubes can be placed on the same rack. Refer to the table below for validated tubes.
CAUTION
Before placing validated closed tubes on the UniCel DxC Systems, check the top of
the cap for any residual blood. Residual blood can contaminate the sample and
affect results. If blood is present, remove it by using a cotton-tipped applicator
stick moistened with DI water. When running in the CTS mode, if tubes off-loaded
from the UniCel DxC Systems have water or droplets of water on the caps, disable
the CTS and contact Beckman Coulter Support Center. Note: Oil on a cap is normal.
Closed Tube Sampling (CTS)
This is an optional feature that allows the system to pierce primary sample tubes. If a CTS assembly
is installed, the operator can enable/disable the CTS assembly. A CTS Tracking database monitors
pierced tubes. If you run a tube again, it will not be pierced again. The sample ID stays in this
database for seven days for a tube that has been pierced.
When a particular Sample ID is manually cleared through Sample Programming, the CTS Tracking
information is also cleared for that Sample ID. If the tube with this particular Sample ID was
previously pierced, reloading this sample tube on any tracking networked instrument will cause it
to be repierced.
NOTE
If a sample ID is cleared at the Host (LIS), the sample ID is NOT cleared in the CTS Tracking database.
•
If a previously pierced sample needs to be rerun, do not clear the Sample ID; use the rerun
procedure.
•
If a Sample ID is manually cleared and reprogrammed for a previously pierced sample, remove
the cap to prevent repiercing.
Table 2.3
Tubes Validated for Closed Tube Sampling
CTS Configuration
Tubes Validated for Cap Piercing
Size
1-Blade Thick CTS
Becton Dickinson VACUTAINER with HEMOGARD
13 × 75mm
13 × 100 mm
16 × 100 mm
Greiner VACUETTE
13 × 75mm
13 × 100 mm
1-Blade Narrow CTS
Sarstedt S-Monovette
75 × 15 mm
a
92 × 15 mm
a. This tube requires a special 5.5 mL rack (PN A18642).
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......