A13914AF
9-124
Maintenance
As-Needed/As-Required Maintenance
Install the Wick Clip Assembly and Blade Assembly
1
Install Wick Clip Assembly (1) onto Wash Tower (2). Make sure both ends of the Wick Clip are
secured. Refer to
2
Carefully pick up the Blade Assembly and orient end points of Blade with Slots (3) on Wick Clip.
Slowly push the Blade Assembly straight down so that the Blade goes through the Wick,
through the Wash Tower, and through the alignment slot at the bottom. Refer to
CAUTION
The points on the end of the Blade are very sharp and have been
exposed to potentially biohazardous fluids. To prevent possible injury or
exposure, avoid the Blade area.
3
Lift up the Blade Assembly again as shown in
. Put the Blade Assembly back on the
plastic tray.
4
Remove the Wick Clip Assembly again as shown in
. Examine the bottom of Wick area
for any loose fibers. Remove any loose fibers with scissors.
NOTE
Loose fibers clog the drain.
5
Repeat Steps 1 and 2 above.
6
Align Blade Holder (2) to screw holes. Install and tighten the two Blade Holder Screws (1) with
a screwdriver. Refer to
7
Install the Cap Piercer Cover (2). Tighten the two cover Screws (1) with a screwdriver. Refer to
.
8
After the Cap Piercer Cover is installed, select
Cancel
on the screen to retain the blade count
since the Blade was not replaced. If
OK
is selected, the system resets the blade count to zero.
9
Select
Exit F10
to exit the
Maintenance
screen.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......