A13914AF
9-19
Maintenance
Monthly Maintenance
9
Figure 9.8
6
Twist off the in-line filter (PN 669212) from the input line and replace.
NOTE
When installing in-line filters, be sure the arrow on the filter is pointing in the direction of reagent
flow.
7
After installation of the filter is complete, reattach reagent input line #30 (that flows into
damper) to the top of the filter.
8
Select:
•
Rgts/Cal
from the menu bar,
•
MC Rgts
,
•
Load F1
.