Maintenance
As-Needed/As-Required Maintenance
3
Clean the inside of the PHOSm cup with an applicator swab moistened with deionized water.
4
Select
OK
when the cup cleaning is complete. The cup primes five times with water.
5
Select
Close
to exit the
Result
pop-up window.
6
Select
Cancel
to exit the
Cup Maintenance
dialog box.
IMPORTANT
The System does not autoprime in
Maintenance
mode. After completing a maintenance
procedure, or when a procedure is aborted, exit the Maintenance menu to resume autoprime. Return
the system to
Standby
.
Inspect the Cuvette Washer Probe
The following procedure is used to inspect the cuvette washer probe. If the cuvette wash station
probe is obstructed by debris (such as glass from a broken cuvette), the DxC may provide inaccurate
results. If any of these steps fail, contact Beckman Coulter Support Center or a local Beckman
Coulter representative to identify and resolve any suspected problems.
1
Remove the Cuvette Wash Station cover. Pull up on the two locking pins. Move the right side of
the cover forward. Carefully lift out the cover.
2
Loosen the two thumbscrews (1) on each side of the cuvette washer and lift off the upper section
of the cuvette wash assembly.
Figure 9.77
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......