A13914AF
9-11
Maintenance
Weekly Maintenance
9
1
Remove the cuvette wash station cover. Pull up on the two locking pins (1). Pivot the right side
of the cover, turning it forward. Lift out the cover, being careful not to disturb the CC sample
mixer.
Figure 9.2
Cuvette Wash Station Cover
2
Loosen the two thumbscrews (1) on each side of the cuvette washer. Lift off the upper section
of the cuvette wash assembly. Pull off the silicone wiper.
Figure 9.3
3
Install new wiper (1) by sliding it onto the bottom of the probe until the under surface of the
wiper is flush with the end of the probe.