A13914AF
8-22
User Defined Reagents
Expanded User Defined Chemistry Feature
On Board Dilution
On Board Dilution allows all samples for the UDR to be diluted prior to analysis. An aliquot of diluent
is placed into a cuvette. Then the Sample is dispensed into the diluent and mixed. On the next cycle,
the diluted sample is aspirated from the first cuvette and dispensed into another cuvette containing
the reagent. The diluent can be either a cartridge of DIL1 (PN 467826) or Component A in the UDR
cartridge.
Enable/Disable On Board Dilution
1
The instrument must be in
Standby
or
Stopped
mode.
2
Select
Setup
from the menu bar.
3
Select
15 User Defined Chemistries
.
4
Select the number to be defined.
OR
Type a number in the
Option Number
field.
5
Select
Define F1
.
6
Select
UDR+ F3
.
7
Go to page 2 of the UDR Setup, and select
Dilute F2
. A new window appears.
8
Type the volume of neat sample to use for the dilution, 3–40 μL.
9
Type the volume of diluent to use for the the dilution, 140–300 μL. The dilution factor
automatically calculates and shows on the screen.
10
Select either check box to define where the diluent will be aspirated from.
UDR Component A
(default)
OR
DIL1 Cartridge
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......