A13914AF
4-1
CHAPTER 4
Reagent Load/Calibration
Reagent Load
Introduction
This chapter describes how to load reagents onto the DxC 800 and DxC 600 systems and calibrate
chemistries using the
Reagent Load/Calibration
screen. The DxC system loads and removes both
cartridge (CC) and modular (MC) chemistries bulk reagents. Reagent information encoded on the
container label can be read by the bar code reader or entered manually from the keyboard.
Cartridge reagents:
•
are stored in wedge-shaped cartridges containing three separate compartments.
•
include a majority of chemistries (excluding the electrolytes).
•
include user-defined reagents.
Modular reagents:
•
Modular reagents are stored inside the lower left compartment in either 500 mL, 1 L, or 2 L
bottles.
•
Other solutions associated with modular reagents include:
—
No Foam
—
CTS Auto-Gloss (used to lubricate CTS blade)
—
Wash Concentrate II
—
Alkaline Buffer
No Foam, CTS Auto-Gloss, and Wash Concentrate II are stored in bottles in the center
Hydropneumatics compartment. The Alkaline Buffer is stored in a bottle within the ISE Module.
Always check reagent status before starting a run. Make sure reagent is loaded for all tests ordered
and that all reagents have sufficient volume to complete the number of tests in the run.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......