A13914AF
3-20
System Setup Options
Special Calculations Definition
Special Calculations Definition
Introduction
The two types of Special Calculations that can be reported with a sample are:
•
Predefined by Beckman Coulter
•
Operator defined
The system can maintain up to 40 special calculations. You can enable or disable the predefined
special calculations, but you cannot modify or delete them from the system. You can define and
modify additional special calculations.
When a special calculation is enabled, the value is reported only if the appropriate chemistries for
the equation are programmed and run for a given sample ID.
The system uses the full precision of the test’s results when performing special calculations. If you
manually calculate, using results that have been rounded, the results will be similar but they will
not match exactly.
From the
Setup
screen, select
12 Special Calculation
to define, edit, delete or view a special
calculation.
Special Calculations may be viewed at any time.
The system must be in
Standby
or Stopped
to define, edit or delete operator-defined Special
Calculations or to enable/disable them.
The default for any Special Calculation is disabled.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......