A13914AF
4-9
Reagent Load/Calibration
Reagent Load
4
Figure 4.8
CC Reagent Information
Load/Unload Cartridge Chemistries (CC)
1
Select
Rgts/Cal
from the menu bar.
2
Select all positions to be loaded or unloaded. The selected positions are highlighted. Use the up
and down arrows to access screens for additional chemistries.
3
Select
Load F1
.
4
To unload reagent:
•
When the screen prompts you to remove the cartridge (1), open the reagent carousel door.
The cartridge to unload is positioned on either the upper or lower reagent carousel. A
blinking bar code reader indicates which carousel position to unload from the upper and
lower reagent carousel. Top positions (2) are 31 to 59. Bottom positions (3) are 1 to 30. Grasp
the cartridge at top and bottom, lift up a little and pull back. Refer to
.
•
As the cartridge is moved in front of the bar code reader, a beep will indicate that the bar
code is read. Check the screen to verify reagent data is removed from the position selected.
If the bar code does not read, manually clear the position. To manually clear, select
Clear F1
.
1.
Reagent Name
2.
Lot Number
3.
Expiration Date
4.
Serial Number
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......