A13914AF
9-39
Maintenance
Two-Month Maintenance
9
5
Refer to
.
Remove the blade and wick assemblies as follows:
•
Hold the blade (1) at the top. Do not touch the points (2) of the blade.
•
Lift up the blade and remove it with the wick assembly at the same time:
—
The wick assembly (3) should be attached to the blade.
—
Do not touch the wick because it contains CTS Auto-Gloss lubricant which is very
slippery and difficult to clean.
•
Discard the old assemblies into a biohazard sharps container.
Figure 9.24
6
Refer to
.
NOTE
A new blade is shipped with a wick attached. The points of the blade are very sharp. The blade
shaft is not sharp.
•
Remove a new blade and wick from its container.
•
Hold it at the top of the blade (1), immediately under the blade clamp as shown in
A011872P.EPS
3
1
2