A13914AF
9-71
Maintenance
As-Needed/As-Required Maintenance
9
7
Unscrew the electrode retainer nut (1) and withdraw the electrode (2) from the flow cell.
Verify that the quad-ring from the removed electrode has not been left in port opening.
Figure 9.50
8
Remove the retainer nut from the old electrode and install on new electrode.
9
Examine the electrode tip. If an air bubble is seen inside the tip, hold the electrode cable near
the end and spin the electrode for a few seconds. The bubble should be pushed back inside the
electrode and remain there away from the tip.
10
Using a lintless tissue, carefully and thoroughly dry the electrode sides and electrode port.
IMPORTANT
Do not touch the electrode tip.