A13914AF
9-62
Maintenance
Six-Month Maintenance
CAUTION
Reaction cup and stir bar cleaning is recommended whenever the sensor is
replaced. Remove the sensor prior to removing stir bar. To prevent damage to the
glucose oxygen sensor membrane tip, do not insert stir bar removal tool,
applicator stick, or any other object into the glucose reaction cup unless the
sensor has been removed.
5
While the old sensor is removed, clean the cup and stir bar. Using the stir bar removal tool,
remove the stir bar. Clean the stir bar and cup with 10% cleaning solution (one part Wash
Concentrate II, nine parts deionized water). Use cotton swabs moistened with the 10% cleaning
solution to clean the cup. Rinse the cup using applicator swabs moistened with deionized water.
Rinse the stir bar with deionized water. Reinstall the stir bar.
6
This step only applies to DxC 800 Systems. For DxC 600 Systems, proceed to Step 8.
Hold the new sensor, connected to the BUNm/UREAm connector panel, in a vertical position
with the membrane tip pointed downward; tap side of sensor body to remove any bubbles from
the membrane tip.
7
This step applies to DxC 800 Systems only. For DxC 600 Systems, proceed to Step 8.
Disconnect the new prepared glucose sensor from the BUNm/UREAm connector panel.
8
Slide the retainer nut over the cable connector onto the new sensor.
9
Align the keyway on the body of the retainer containing the sensor with the key in the sensor
port and install the sensor assembly. Finger-tighten the retainer nut.
10
Reconnect the pin lead into the glucose connector panel by aligning the latch on the connector
with the notch on the connector panel. Push the pin lead into the connector panel until the
latch clicks.
11
Select
OK
in the Maintenance pop-up confirmation window when replacement is complete. The
Glucose cup is primed five times with water.
While module is priming:
•
Observe the reagent line and Glucose reaction cup to determine there are no leaks. If any
leaks are noticed, correct the problem.
•
Observe the stir bar to verify that it is rotating properly.
12
Select
Close
when priming has completed to exit the
Result
pop-up window.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......