A13
9
14
AF
12
-1
5
T
roubl
es
hoo
tin
g
C
a
libr
a
ti
on and
Res
u
lt
E
rr
o
rs
Co
m
m
o
n
Er
ro
r Me
ss
a
g
e
s a
n
d C
o
rr
e
ct
ive
A
ct
io
n
s
12
0x02620025
2
CC Sample Syringe
Motion Error
CAU:
CC Sample Syringe
Motion Error
Message ID:
0x02620025
Subsystem:
RASched
Action Name:
action name
This error occurs when there is an error
within the Sample Syringe Subsystem.
This includes both the Shear valve Motor
and the Lead Screw Motor.
•
Make sure syringe and plunger are properly
seated at top and base of syringe.
•
Check for obstruction of syringe movement.
•
Perform CC Syringe Home Alignment.
•
If the error continues, contact Beckman Coulter
Support Center for additional information.
0x02620033
4
10 Cuvettes Have failed
Water Blank
CAU:
10 Cuvettes Have Failed
Water Blank
Message ID:
0x02620033
Subsystem:
RaSched
This error can be flagged for the
following:
•
A cuvette's wavelength absorbance
is over 350 mA (Absorbance) for all
wavelengths between 340 to 700
nanometers.
•
A cuvette's wavelength absorbance
is over 200 mA (Absorbance) for the
940 nanometer wavelength (PRO
System only).
•
PSM Error.
•
LPIA Failure.
•
Remove wash station head and inspect probes
for blockage and wiper for wear. Refer to
for removal of Wash
Station.
•
Check wash station wiper alignment.
•
Check wash station alignment.
•
Remove reaction carousel and check cuvette
cleanliness.
Non-resolution of problem may indicate photometer
issues. Contact Beckman Coulter Support Center for
additional information.
0x02620034
4
CC Reagent probe
disabled due to level
sense errors
CAU:
CC Reagent Probe
Disabled Due To level Sense
Errors
Message ID:
0x02620034
Subsystem:
RASched
Probe:
probe
Multiple reagent cartridges have level
sense errors.
•
the system.
•
Check reagent cartridges for bubbles.
•
Perform Reagent Probe to Cartridge and
Reagent Probe Height in Cartridge Alignments.
Refer to step 3 of the
procedure.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......