A13914AF
11-3
Utilities
Prime
11
Stopping a Prime
When a Prime has started, you can stop the Prime before the selected number of replicates has
completed.
From any of the prime screens, select
Stop Prime
.
The replicate that is currently processing will complete and any remaining primes will be canceled.
Priming All Subsystems
If the entire group of subsystems, such as all Cartridge Chemistry (CC) subsystems, needs to be
primed, it is easiest to simply select the group of subsystems for prime instead of having to select
each individual subsystem. Follow the steps below to prime a group of subsystems.
1
Select
Utils
from the menu bar.
2
Select
1 Prime
.
OR
Type
1
in the
Option Number
field and press
(
Enter
)
.
3
Select one or a combination of the following check boxes:
All CC Subsystems
All Hydropneumatic Subsystems
All MC Subsystems
Prime DxC
CTS Auto-Gloss (CTS System only)
CTS Blade Wash (CTS System only)
.
4
Type the desired number of primes (1–99) in the
Number of primes to repeat
field. This does not
apply for Hydro primes.
5
Select
Start Prime
. The system status changes to
Priming
.
6
Upon completion of the requested prime, the number of primes remaining will be blank and the
system status returns to
Standby
.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......