A13
9
14
AF
12
-3
7
T
roubl
es
hoo
tin
g
C
a
libr
a
ti
on and
Res
u
lt
E
rr
o
rs
Co
m
m
o
n
Er
ro
r Me
ss
a
g
e
s a
n
d C
o
rr
e
ct
ive
A
ct
io
n
s
12
0x13A000CE
10
Samples in rack #
<rack number> have
barcodes longer than
15 characters. Rack will
be unloaded from
<rack location>
Position Sample ID
1 ID too long
2 ID too long
3 ID too long
4 ID too long
WLM:
Rack contains samples
with barcodes > 15 characters.
Rack:
<rack #>
[1:<cup1 ID>]
[2:<cup2 ID>]
[3:<cup3 ID>]
[4:<cup4 ID>]
One or more samples were loaded with
barcodes longer than 15 characters.
•
Check barcodes on samples in rack.
•
Reload rack.
•
If the error continues, contact Beckman Coulter
Support Center for additional information.
0x13A000E4
10
Module: <module
name> is disabled due
to a stirrer motor error.
All module functioning
has now ceased. Call
Beckman Coulter
Technical Support.
WLM:
Stirrer motor error -
module disabled: <module
name> (<file>:<line>)
If a motor speed error occurs, the
following symptoms are exhibited:
•
With
Albumin (ALBm)
,
Blood Urea
Nitrogen/Urea (BUNm/UREAm)
,
Glucose (GLUCm)
, and
Total
Protein (TPm)
, results are
suppressed with an Initial Rate Low/
High Error Flag. When three
consecutive results are suppressed,
the Cup Module is disabled.
•
With
Creatinine (CREm)
,
suppressed results are reported with
a Reaction Noise Error Flag. The
CREm Cup Module is
not
disabled; it
continues to run and reports
suppressed results with an error flag.
•
With
Phosphorous (PHOSm)
, low
results are reported. The PHOSm
Cup Module is not disabled; it
continues to run and reports low
results.
•
Contact your local service representative to
schedule an inspection of the stirrer motor.
DO
NOT use the cup module until your service
representative has inspected and/or replaced
the stirrer motor.
•
Review the results reported just prior to the
occurrence of these symptoms and respond in
accordance with your laboratory policies and
procedures.
•
From Instrument Commands, disable the
appropriate cup module.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......