0x057A0008
4
Rack Out of Range
CAU:
Invalid Rack ID
Message ID:
0x057A0008
Subsystem:
SRShttl
The bar code reader was unable to
identify the rack within a specified range
(1-999) during a load or unload of the
rack.
•
If this occurs on a rack load:
—
Check that the bar code label is on the rack
and facing to the right.
—
Check the condition of the bar code label on
the rack.
•
If this occurs on a rack unload, perform
Unload
All
. Refer to
.
•
If the error continues, compare the racks shown
on the Main screen with the racks present on the
Sample Carousel to make sure the rack is
physically present.
—
If on Main screen, but the rack is not
physically present, reboot system. Refer to
—
If on Main screen, and the rack is physically
present, shuttle latch may not have hooked
the rack properly. Contact Beckman Coulter
Support Center for additional information.
•
If this error occurred during a rack unload, check
the results for the samples in the misread rack.
0x057A0009
4
Duplicate Rack ID
rack
#
CAU:
Duplicate Rack ID
Message ID:
0x057A0009
Subsystem:
SRShttl
Rack ID:
rack #
This error occurs during a load when a
rack ID is read by the bar code reader
and an identical rack ID is already
present on the system. The system
cannot use duplicate ID racks
simultaneously.
•
Compare the number of the rack that was
rejected with the racks currently on the analyzer
by looking at the Main Screen.
•
Check the actual bar code that is on the rack
correlates with the number that the system is
rejecting.
•
Switch samples to a rack with a different ID or
wait until the samples in the previous rack has
completed testing and has been offloaded.
•
If the Main screen shows the Rack, but it is
physically not on the instrument, use the Unload
All feature to clear the sample wheel. Refer to
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......