A13914AF
9-84
Maintenance
As-Needed/As-Required Maintenance
13
Using the syringe filled with 10% bleach solution, connect the syringe tip over the flow cell exit
port #23 (1).
Figure 9.66
14
Flush the 10% bleach solution through the flow cell exit port. This flushes the area between the
valve port and flow cell exit port. Allow the bleach solution to sit for one minute.
The one-minute waiting period applies only to the 10% bleach solution, and not to the following
deionized water or 70% isopropyl alcohol.
15
Repeat the flushing procedure through the flow cell exit port using deionized water.
16
Repeat the flushing procedure through the flow cell exit port using 70% isopropyl alcohol.
17
Repeat the flushing procedure through the flow cell exit port using deionized water.
18
Thoroughly dry the EIC and attached solenoid valve using a dry lintless tissue. Wipe the drip
tray with deionized water and lintless tissue. Thoroughly dry the drip tray.