A13914AF
4-29
Reagent Load/Calibration
Slope Offset Adjustment
4
5
Move the cursor to the set point to be modified and type the desired value.
•
If a previously modified set point is to be returned to the original default value for that
calibrator or chemistry, select desired chemistry, move cursor to the applicable set point
and select
Restore Defaults
.
•
The default values have more significant figures than shown on the screen. Therefore,
select
Restore Defaults
, rather than typing the default values.
6
Select
Close
to exit the screen.
•
Calibration status changes to
Cal Required
.
•
A flag indicating that specific set points have been modified will appear in the Instrument
Code section on the results report.
Slope Offset Adjustment
Introduction
The Slope Offset Adjustment option allows adjustment of the slope (m) or the y-intercept offset (b)
of the regression equation used for results calculations. Modifications are made on a per chemistry
basis. The default value for the slope is equal to "1" and the default value for offset is equal to "0."
IMPORTANT
Adjusting the slope and offset for a chemistry may result in a mismatch of normal range values
previously established.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......