A13914AF
9-21
Maintenance
Monthly Maintenance
9
16
Calibrate the CO
2
chemistry.
Adjust the Fluid Level of the Alkaline Buffer Damper Assembly
1
Visually inspect the damper to verify that the fluid level rests between the MIN and MAX lines,
approximately one-third to one-half full.
If the fluid level is below the MIN line, proceed to Step 2.
If fluid level is above the MAX line, go to Step 8.
2
Remove the alkaline damper (1) from the holding clip and place the damper in a horizontal
position.
Figure 9.10
NOTE
The damper should be positioned so that the output line #29 (2)
, leading from the
damper to the flow cell, is oriented on top.
3
Prime alkaline buffer five times or as many times as necessary to fill the damper one-half full.
Access alkaline buffer priming.
•
Select:
—
Utils
from the menu bar,
—
1 Prime
,
—
MC F4
,
—
ISE CO
2
Alkaline Buffer
.
A015938P.EPS
1
2