A13914AF
9-32
Maintenance
Monthly Maintenance
Replace CTS Blade (1-Blade Narrow CTS Option)
. Replace the Cap Piercer blade approximately once a month. If the system runs
>500 samples per day, replace the blade more frequently. The blade count records the number of
caps pierced by the blade. Refer to
.
CAUTION
The Cap Piercer blade is very sharp and has been exposed to potentially
biohazardous fluids. To prevent injury or exposure, do not touch the points of the
blade and wear gloves.
Figure 9.18
Cap Piercer Blade
Materials Required:
•
Medium straight-edge screwdriver
Replacement Procedure
1
To access this procedure,
•
Select:
—
Utils
from the menu bar,
—
2 Maintenance
,
—
1 CTS Blade Replacement.
•
To start the procedure select
Continue
.
OR
•
To exit this screen select
Cancel
.
2
The instrument prepares for blade replacement, then the Maintenance pop-up confirmation
window appears.
A014
8
4
8
P.EPS