A13914AF
9-36
Maintenance
Two-Month Maintenance
Two-Month Maintenance
Replace CTS Blade/Wick (1-Blade Thick CTS Option)
Replace CTS Blade/Wick (1-Blade Thick CTS Option)
Refer to
. Replace the Cap Piercer blade and its CTS Auto-Gloss wick approximately every
two months. If the system runs >500 samples per day, replace the blade and wick more frequently.
The blade count records the number of caps pierced by the blade. Refer to
.
CAUTION
The Cap Piercer blade is very sharp and has been exposed to potentially
biohazardous fluids. To prevent injury or exposure, do not touch the points of the
blade and wear gloves.
Figure 9.21
Cap Piercer Blade with CTS Auto-Gloss Wick
Materials Required:
•
Medium Phillips screwdriver
•
Small Phillips screwdriver
•
Medium straight-edge screwdriver
1.
Blade Clamp
2.
Blade
3.
Points
4.
Wick Clip
5.
Wick (in clip housing)
A007367P.EPS
1
2
3
4
5