A13914AF
9-41
Maintenance
Two-Month Maintenance
9
9
Refer to
.
Install the Cap Piercer cover and tighten the two captive Cap Piercer cover screws (1).
10
Select
OK
.
•
The system sets the blade count to zero.
•
The Cap Piercer carriage moves up to
Home
position.
•
The system primes the wick with CTS Auto-Gloss lubricant.
Change/Clean Air Filters
IMPORTANT
To minimize system down time, it is recommended to keep on hand a clean set of air filters.
Dirty filters can then be cleaned at a convenient time.
1
Open all lower compartment doors (left, middle and right).
2
Pull up and remove each of the compartment air filters. There are two different sizes of air
filters. They are located over vent areas on the inner side of each door.
3
Vacuum or brush each filter thoroughly to remove dust. If excessively dirty, rinse with
deionized water. Vigorously shake residual water from filters. Allow the filter to air dry
completely.
CAUTION
Do not place a damp filter back on the system. Residual moisture may cause
damage to the system. You may find it helpful to order a second set of filters to
use as an alternate. The DxC 600 requires two of PN A08879 and one each of PN
A08851 and 466357. The DxC 800 requires three of PN A08879 and one of PN
466375.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......