A13914AF
9-37
Maintenance
Two-Month Maintenance
9
Replacement Procedure
1
To access this procedure, select:
•
Utils
from the menu bar
•
2 Maintenance
•
1 CTS Blade Replacement
.
To start the procedure, select
Continue
.
OR
To exit this screen, select
Cancel
.
2
The instrument prepares for blade and wick replacement, then the
Maintenance
pop-up
confirmation window appears.
CAUTION
The points of the Blade are very sharp and extend below the Wash Tower. Avoid
the bottom of the Wash Tower.
CAUTION
DO NOT select OK when the cover is removed. If you select OK the Cap Piercer
Carriage moves up and causes a pinch hazard.
3
Fully loosen the two captive Cap Piercer cover screws (1) at the base of the Cap Piercer cover.
Lift and remove the cover.
Figure 9.22
A014462P.EPS
1